TLDR: LLMs uplift novices in setting up wetlab equipment and can sort of move things around, but probably still need expert help
There's a lot of talk recently about this chart and AI pretraining algorithmic progress going superexponential, but even the best algorithm needs training data to be useful. One of the most data scarce areas is biology, which makes sense because your RL environment verifier is a wetlab and rats instead of unit tests and the Lean 4 compiler. An interesting and IMO understudied aspect of RSI is AI increasing biological data generation, both from new assays and reducing the barrier to entry to wetlab work.
Traditionally, everyone from startups to big pharma outsourced their wetlab work to Contract Research Organizations (CROs) such as IQVIA, WuXi AppTec, Pharmaron, Charles River, etc. You come up with a set of experiments you want to run, pay 6 to 7 figures, wait a few months for the reagents to ship and cells to grow and you get your data.
Recently, everyone and their grandmother is going into biology. Startups like Adaptyv and Onava offer protein binding assays starting from $99 a binder with lead times as short as 3 weeks, while even high schoolers are starting their own CROs. How much will this outpouring of talent and capital affect the amount of biological data available? What kinds of wetlab tasks can AI accelerate today?
I wanted to personally test this with a simple experiment: Can AI help a wetlab novice (me) set up a Biolayer Interferometry Machine and conduct binding affinity assays?
Binding Affinity
In biology, proteins do stuff by sticking to each other. This is determined by their shape, which is determined by their sequence of amino acids. Therapeutically, you might want to design a drug to bind to a specific protein, to stop it from doing its thing, or deliver something to a certain cell with a specific receptor. For example, antibodies are proteins with highly specific binding affinity, produced by B cells (a type of cell in the immune system).
Antibody based drugs
Examples of proteins binding in non-receptor tyrosine kinase signaling pathways. Notice how they kind of fit together like lego bricks. Source
Previously, you would find these antibodies and binders by brute force (library screening of millions of genes), or naturally find them in animals like rabbits or camels. Now there are many binding affinity models such as Boltz, and generative protein models like BoltzGen and RFdiffusion, allowing you to vastly decrease the search space in silico before moving to the wetlab.
Biolayer Interferometry
The science and entrepreneurial story of BLI is pretty awesome. Biolayer interferometry was invented by Dr Hong Tan in 2001 and commercialized by his company ForteBio. ForteBio was eventually bought by Pall Bio, then Sartorius, which still sells the machines today. Dr Tan then spun out a new company, GatorBio, which offers next generation BLI, and recently won a patent dispute against Sartorius and his own former patents. Basically, all the BLI machines in the world are from this guy's companies (it's incredible how much impact a scientist can have if you give them VC money).
BLI works by shining light through thin glass needles coated with sensor molecules and dipping them into solutions of protein and then analyte (the thing you want to test binding).
SA (Streptavidin) biosensor, with biotinylated protein. When making your protein, you tag it with vitamin B7 (biotin) which streptavidin has extremely high binding affinity for. The analyte in this case is the antibodies in step 3. Source
What the sensors look like in the machine. The instrument head moves and dips the sensors in different rows for certain time periods. For instance, your protein might be in column 1, baseline in column 2, analyte in column 3, and regeneration buffers in 4 and 5
There are two boundaries where the medium changes: One from glass to sensor molecules/bound proteins, and one from the molecules to ambient air or solution. As binding occurs at the needle tip, this changes the length of the biomolecule layer. This change shifts the interference pattern slightly, which is picked up by a spectrophotometer. You'll often see BLI charts with nanometer response (shift) on the y-axis and time on the x-axis, giving you raw data to calculate both binding affinity and kinetics.
There are many different types of coatings you can have, from Anti-HIS to Streptavidin, depending on the type of protein and tag.
BLI Math
Binding affinity is typically measured by the dissociation constant , which is the analyte concentration at which half of protein sites are occupied at equilibrium.
Let's say you have a protein P on the sensor tip and analyte A in solution, and they reversibly bind to form a complex PA.
The forward rate of reaction is and reverse rate is , where [] is concentrations in M (mol/L) and and are experimentally determined constants. P is immobilized on the BLI sensor tip, so [P] could also be interpreted as mol/unit area. has units and has units , since reaction rates have units M/time (usually seconds). Intuitively, does not depend on concentration of another analyte, only the concentration of the bound complex.
The rate of bound complex formation is
At equilibrium, , the rate of proteins sticking to the sensor equals the rate of proteins falling off, and is defined as this ratio
At 50% occupancy, [P] = [PA], so is an analyte concentration [A]. can be interpreted as the concentration of analyte needed to fill up half the reaction sites at equilibrium.
Notice the ratio of is a ratio of free sites divided by bound sites. You can't directly measure the concentration of free and bound sites to give you . But in BLI land, you can estimate it in terms of nm response units, and . Let's say you do a dilution series of analyte. When the curves level off, you record the following responses.
Starting analyte concentration in the well [A]
10 nM
0.45 nm
100 nM
0.82 nm
1,000 nM
0.89 nm
It might look something like this:
From this you can estimate that if the sensor were to be saturated, you would get around 0.9 nm response, so .
Now, you have an estimate of free binding sites over bound binding sites for each analyte concentration.
You can then calculate the from each row of the table, since you know the , , and [A]. For example in the 10 nM row:
Starting analyte concentration in the well [A]
10 nM
0.45 nm
10 nM
100 nM
0.82 nm
9.76 nM
1,000 nM
0.89 nm
11.24 nM
And we can see that based on this data, is around 10 nM.
It is also possible to estimate from a single curve, for example the one with 10 nM analyte concentration.
In the association phase (sensors dipped in analyte) the response over time can be modeled by
Where .
Intuitively, depends on both the rate of binding and rate of dissociation.
Substituting R(34.7)=0.225, we can calculate
Now we calculate from the dissociation curve (no new binding is happening, the tips are moved to assay buffer with no analyte). We model dissociation as
and substitute R(69.3) = 0.225
Calculate by subtracting from and dividing by [A]
Finally calculate
Hardware
New BLI machines such as the Sartorius Octet cost around $250,000 for an 8-channel configuration. You can go far cheaper by buying a used, older model on ebay. I found an extremely scuffed one for $1,819 (and free shipping!)
But let's do even better than that. Government and private auction sites like Govdeals.com and Heritage Global Partners routinely auction off old surplus equipment from universities and biotechs. Chat was very helpful in scanning all upcoming auctions, making a spreadsheet, and doing market research for used prices. Here's one from Amgen I got for $550.
Unfortunately, after taxes and "Buyer's Premium" it ended up costing $724.86.
It also came with absolutely no cables or power supply. The power supply I found on ebay for $199, and the wires for $60 from SFCable.
An extremely cooked wiring setup for RD9 instrument control, USB A ejector and spectrophotometer out, and USB B shaker/heater control
Software
At first, I tried getting GPT 6 Astra, Claude Sonnet 5.5 (the other Claudes refused), and Kimi K3 to read the USB serial ports and reconstruct the API that way. And this sort of worked! The models were able to move the instrument head around, shake the heated shaker, and run protocols.
Kimi K3 in particular was extremely excited to figure out how the shaker worked
They also figured out the instrument control was from Galil and the Spectrophotometers from Ocean Optics
I used GPT 6 Astra to vibecode a Mac app to visualize the spectra, manually move the instrument, and run protocols.
Instrument Controls. Chat was very safe and required that the lid be closed, the motion path be clear, and a human operator be nearby
Raw spectrophotometer readings
Here is the output of one protocol, quantitation of His-tagged sfGFP (just a single protein, not two binding to each other) and regeneration (washing off the bound protein to reuse the sensor).
HIS1K Biosensor measuring His-tagged sfGFP. Two loading and dissociation cycles with 3 regeneration cycles in between.
Unfortunately, chat didn't know how to set reference dark and light files or do live streaming of nanometer response vs time graph. I assume this is because it didn't know how to convert raw spectrophotometer counts to nm response. It also made a loud mechanical screeching noise every time it moved the instrument head around, which wasn't great.
Expert Help
Since that didn't quite work out, I had no choice but to ask dreaded customer support for help.
The entire bio hardware and lab automation industry is known for extremely walled off and proprietary software. It's like Apple, if there were a hundred different Apples and instead of the most beautiful product design they did the opposite.
If you call Beckman Coulter and ask them for software to run their Biomek liquid handler, they will refuse to send the software over email, website portal, or any method over the internet. They will instead send a field engineer with a laptop with the software pre-installed and sell you both the laptop and consulting services for around $27,000.
Luckily, Sartorius customer support was far nicer and gave me a copy of the software for free, since they were no longer supporting the ForteBio line of instruments.
Ugly but functional (and free!) BLI data acquisition software from Sartorius
Even better, they cc'd a field application specialist (forward deployed engineer) to help out, since normally these machines are installed by expert technicians. "Robert" from Sartorius came by and was like ye it's not supposed to make that noise, and I was like ye I have no idea what I'm doing. So Robert invited me to the Sartorius showroom in Fremont for a demo (and even drove me there from SF). We entered this large, white, beautifully stocked lab with tons of gleaming, strange equipment, and I had the vague feeling of being shown around a car dealership (complete with the same general level of prices).
The Sartorius Demo Lab. Not only did they have BLI machines in 1, 2, 4, 8, 16, and 96 channel configurations, they also had bioreactors, surface plasmon resonance machines, robotic cell sorters, and more!
We spent about 8 hours on a binding kinetics protocol (human IgG + biotinylated protein A) and a quantitation experiment with some sfGFP I brought. Robert showed me everything from prepping the machine (we used an Octet R8e), to preparing the reagents and dilution series, to running the data analysis software and fitting data to theoretical curves.
Plate definition: B is assay buffer (phosphate-buffered saline solution), L is loading (protein A), the purple is the analyte (human IgG). R is acidic pH 1.5 10 mM glycine for regeneration, and N is neutralization, which is the same as assay buffer. To remove human IgG from the sensor and reuse, we do 3 regeneration cycles by quickly dipping the sensors in R and then N for 5 seconds each.
There were so many things to know, including:
The machine should stay on continuously unless you aren't using it for months, because turning it on and off frequently lowers the tungsten bulb lifespan
The machine should be used with only the Greiner 655209 well plate since it has the correct optical characteristics
A dilution series with at least 3 concentrations is needed to calculate binding kinetics accurately
You must experimentally determine the optimal concentrations for each protein. Ideally, the amount of protein in solution binds one to one to the sensor tip coating, and if there is too high of a concentration there can be steric hindrance and other data quality issues.
The sensors must be hydrated at least 10 minutes in assay buffer to remove the protective sucrose coating. If they are expired according to the package, soak for 20 minutes
It was honestly the most wonderful sales experience I ever had (we even got lunch at a nearby noodle place).
Here are some pretty curves we got at the end.
SA sensor for biotinylated protein A + human IgG. The first upward curve from 60-120 seconds is the loading of the protein A onto the sensor. This is followed by 3 regeneration cycles and a baseline soak. 210-390 is the association phase, when we dip in the human IgG solution. 390-570 is dissociation, where the bound IgG is soaked in fresh assay buffer and slowly leaves the sensor.
HIS1K sensor for His-tagged sfGFP. We didn't do regeneration and only tested loading, since it was just one protein.
Afterward, Robert gave me about $3,000 worth of biosensors, well plates, and sample reagents to troubleshoot my old machine back at the lab.
I imagine this is how children feel on Halloween
Running the Machine
Armed with new software and goodies, I tried replicating the biotinylated protein A + human IgG assay. The machine moved much more smoothly and didn't make wild noises.
There was one more problem. The machine needed initial calibration with dark (lamp off) and light (sensors in baseline) reference measurements. To set these, you have to open the software in diagnostic mode, which entails holding Ctrl + Shift while clicking the shortcut (Chat figured this out but got the timing wrong, telling me to click and then hit Ctrl+Shift on startup).
This opens diagnostic mode, which has absolutely ZERO documentation in the user guide or online (either because the machine is too old or this tacit knowledge is typically only known by a field application specialist).
What do I even do here
After a ton of back and forth with chat and hitting random buttons to observe their effects, I eventually figured out what most of the things did.
Snap: Instantaneous spectrophotometer reading
Start/Stop Scan: Take spectrophotometer readings over time, optionally graph them on the top right chart
Save Dark: Save readings when lamp is off
Save Reference: Save readings in baseline assay buffer
Save Binding: Save readings when binding to analyte.
Auto Gain (Now!): Based on the lighting and spectrophotometer intensity, change the integration time (essentially exposure time). Typically this is 1-3 ms, with higher times letting in more light
Set Integration: Manually set integration time
Go Sample x: Goes to a column number x and dips the sensors
Pick Sensor x: Goes to sensor tray column x and loads the sensors
I placed two sensor tips in the sensor tray and hydrated them for 20 minutes. I noticed they were in the wrong position, and moved them to the correct positions, which didn't have assay buffer under them. This will be important in a second.
Prepping buffers, reagents, and soaking the two sensors on the right in assay buffer. There is another one of the black Greiner 655209 well plates under the green holder in the sensor tray. However, not all of these wells are filled, and the ones the protocol used to calibrate integration time were empty.
I loaded up the .fmf method file for the protocol, checked everything was good, and hit GO. The machine started the protocol, picked up the sensors, but the second it went to the first well, the sensor failed with errors (CCD signal too low) and the spectrophotometers showed zero. This was weird, because in diagnostic mode the sensors were working fine.
Data error on sensors A and B
At first I thought the calibration was off, and tried resetting the references, trying different channels, setting different INI file configs. I tried changing the integration time to 3 ms. But I noticed on assay start, it would automatically revert to ~1 ms! It turned out the sensors sitting in empty wells reflected more light than if they were sitting in assay buffer (which they were supposed to I guess) and this causes auto gain to automatically decrease integration time at the beginning of each protocol.
Notice in Instrument Status on the bottom left Sensor 1: Integration Time is set to 1.65 ms and Sensor 2 to 1.78 ms
I pipetted assay buffer under the sensors in the tray, and the sensors started working!
Biotinylated Protein A + human IgG. Red is control (assay buffer) and blue is human IgG. The nanometer response is inverted for some reason
I also tried regenerating the sensor tip but not very successfully, since at 370 seconds the curves don't drop back to 0. Here are the final curves:
Biotinylated Protein A + human IgG, two cycles + regeneration. Red is control (assay buffer) and blue is human IgG. The nanometer response is inverted for some reason
I asked chat to calculate the , , and from the data, and is apparently 59 times greater than published Octet values (indicating weaker binding), but close to another published study with porous silica.
Overall it took about $1000 and 1 month of part-time work to get from nothing to .
Limitations
This is an n of 1 case study with just 1 idiot in 1 wetlab. Your results may vary
I didn't try making my own proteins yet, will probably do cell-free expression for that (the sfGFP was from a friend). This will likely take even longer
I didn't do a control of someone setting up a BLI machine from scratch without LLMs
Conclusion
What were the LLMs good at today?
Scanning auction sites
Designing experimental protocols and choosing sensor types
Driving lab equipment from USB serial ports
Doing the math to calculate kinetic constants from raw data (.frd files)
What could have made them more helpful?
Ability to reverse engineer lab equipment software
Better scientific multimodality and physical intelligence. Having eyes, ears, and hands
More tacit bio knowledge, especially from current wetlab technicians/field application specialists. Like the ability to replace anything Robert from Sartorius could tell me
Even though they are underwhelming at present, I expect LLMs to become far more useful and powerful in the wetlab in the coming year. There are many startups like Transfyr and C5R that offer to sell multimodal wetlab data to the labs. LLMs can already (slowly) drive robot arms. Just like how LLMs could barely do math a year or so ago and are now superhuman, with the correct data and training environments the same should happen for biology.
TLDR: LLMs uplift novices in setting up wetlab equipment and can sort of move things around, but probably still need expert help
There's a lot of talk recently about this chart and AI pretraining algorithmic progress going superexponential, but even the best algorithm needs training data to be useful. One of the most data scarce areas is biology, which makes sense because your RL environment verifier is a wetlab and rats instead of unit tests and the Lean 4 compiler. An interesting and IMO understudied aspect of RSI is AI increasing biological data generation, both from new assays and reducing the barrier to entry to wetlab work.
Traditionally, everyone from startups to big pharma outsourced their wetlab work to Contract Research Organizations (CROs) such as IQVIA, WuXi AppTec, Pharmaron, Charles River, etc. You come up with a set of experiments you want to run, pay 6 to 7 figures, wait a few months for the reagents to ship and cells to grow and you get your data.
Recently, everyone and their grandmother is going into biology. Startups like Adaptyv and Onava offer protein binding assays starting from $99 a binder with lead times as short as 3 weeks, while even high schoolers are starting their own CROs. How much will this outpouring of talent and capital affect the amount of biological data available? What kinds of wetlab tasks can AI accelerate today?
I wanted to personally test this with a simple experiment: Can AI help a wetlab novice (me) set up a Biolayer Interferometry Machine and conduct binding affinity assays?
Binding Affinity
In biology, proteins do stuff by sticking to each other. This is determined by their shape, which is determined by their sequence of amino acids. Therapeutically, you might want to design a drug to bind to a specific protein, to stop it from doing its thing, or deliver something to a certain cell with a specific receptor. For example, antibodies are proteins with highly specific binding affinity, produced by B cells (a type of cell in the immune system).
Antibody based drugs
Examples of proteins binding in non-receptor tyrosine kinase signaling pathways. Notice how they kind of fit together like lego bricks. Source
Previously, you would find these antibodies and binders by brute force (library screening of millions of genes), or naturally find them in animals like rabbits or camels. Now there are many binding affinity models such as Boltz, and generative protein models like BoltzGen and RFdiffusion, allowing you to vastly decrease the search space in silico before moving to the wetlab.
Biolayer Interferometry
The science and entrepreneurial story of BLI is pretty awesome. Biolayer interferometry was invented by Dr Hong Tan in 2001 and commercialized by his company ForteBio. ForteBio was eventually bought by Pall Bio, then Sartorius, which still sells the machines today. Dr Tan then spun out a new company, GatorBio, which offers next generation BLI, and recently won a patent dispute against Sartorius and his own former patents. Basically, all the BLI machines in the world are from this guy's companies (it's incredible how much impact a scientist can have if you give them VC money).
BLI works by shining light through thin glass needles coated with sensor molecules and dipping them into solutions of protein and then analyte (the thing you want to test binding).
SA (Streptavidin) biosensor, with biotinylated protein. When making your protein, you tag it with vitamin B7 (biotin) which streptavidin has extremely high binding affinity for. The analyte in this case is the antibodies in step 3. Source
What the sensors look like in the machine. The instrument head moves and dips the sensors in different rows for certain time periods. For instance, your protein might be in column 1, baseline in column 2, analyte in column 3, and regeneration buffers in 4 and 5
There are two boundaries where the medium changes: One from glass to sensor molecules/bound proteins, and one from the molecules to ambient air or solution. As binding occurs at the needle tip, this changes the length of the biomolecule layer. This change shifts the interference pattern slightly, which is picked up by a spectrophotometer. You'll often see BLI charts with nanometer response (shift) on the y-axis and time on the x-axis, giving you raw data to calculate both binding affinity and kinetics.
There are many different types of coatings you can have, from Anti-HIS to Streptavidin, depending on the type of protein and tag.
BLI Math
Binding affinity is typically measured by the dissociation constant , which is the analyte concentration at which half of protein sites are occupied at equilibrium.
Let's say you have a protein P on the sensor tip and analyte A in solution, and they reversibly bind to form a complex PA.
The forward rate of reaction is and reverse rate is , where [] is concentrations in M (mol/L) and and are experimentally determined constants. P is immobilized on the BLI sensor tip, so [P] could also be interpreted as mol/unit area. has units and has units , since reaction rates have units M/time (usually seconds). Intuitively, does not depend on concentration of another analyte, only the concentration of the bound complex.
The rate of bound complex formation is
At equilibrium, , the rate of proteins sticking to the sensor equals the rate of proteins falling off, and is defined as this ratio
At 50% occupancy, [P] = [PA], so is an analyte concentration [A]. can be interpreted as the concentration of analyte needed to fill up half the reaction sites at equilibrium.
Notice the ratio of is a ratio of free sites divided by bound sites. You can't directly measure the concentration of free and bound sites to give you . But in BLI land, you can estimate it in terms of nm response units, and . Let's say you do a dilution series of analyte. When the curves level off, you record the following responses.
Starting analyte concentration in the well [A]
10 nM
0.45 nm
100 nM
0.82 nm
1,000 nM
0.89 nm
It might look something like this:
From this you can estimate that if the sensor were to be saturated, you would get around 0.9 nm response, so .
Now, you have an estimate of free binding sites over bound binding sites for each analyte concentration.
You can then calculate the from each row of the table, since you know the , , and [A]. For example in the 10 nM row:
Starting analyte concentration in the well [A]
10 nM
0.45 nm
10 nM
100 nM
0.82 nm
9.76 nM
1,000 nM
0.89 nm
11.24 nM
And we can see that based on this data, is around 10 nM.
It is also possible to estimate from a single curve, for example the one with 10 nM analyte concentration.
In the association phase (sensors dipped in analyte) the response over time can be modeled by
Where .
Intuitively, depends on both the rate of binding and rate of dissociation.
Substituting R(34.7)=0.225, we can calculate
Now we calculate from the dissociation curve (no new binding is happening, the tips are moved to assay buffer with no analyte). We model dissociation as
and substitute R(69.3) = 0.225
Calculate by subtracting from and dividing by [A]
Finally calculate
Hardware
New BLI machines such as the Sartorius Octet cost around $250,000 for an 8-channel configuration. You can go far cheaper by buying a used, older model on ebay. I found an extremely scuffed one for $1,819 (and free shipping!)
But let's do even better than that. Government and private auction sites like Govdeals.com and Heritage Global Partners routinely auction off old surplus equipment from universities and biotechs. Chat was very helpful in scanning all upcoming auctions, making a spreadsheet, and doing market research for used prices. Here's one from Amgen I got for $550.
Unfortunately, after taxes and "Buyer's Premium" it ended up costing $724.86.
It also came with absolutely no cables or power supply. The power supply I found on ebay for $199, and the wires for $60 from SFCable.
An extremely cooked wiring setup for RD9 instrument control, USB A ejector and spectrophotometer out, and USB B shaker/heater control
Software
At first, I tried getting GPT 6 Astra, Claude Sonnet 5.5 (the other Claudes refused), and Kimi K3 to read the USB serial ports and reconstruct the API that way. And this sort of worked! The models were able to move the instrument head around, shake the heated shaker, and run protocols.
Kimi K3 in particular was extremely excited to figure out how the shaker worked
They also figured out the instrument control was from Galil and the Spectrophotometers from Ocean Optics
I used GPT 6 Astra to vibecode a Mac app to visualize the spectra, manually move the instrument, and run protocols.
Instrument Controls. Chat was very safe and required that the lid be closed, the motion path be clear, and a human operator be nearby
Raw spectrophotometer readings
Here is the output of one protocol, quantitation of His-tagged sfGFP (just a single protein, not two binding to each other) and regeneration (washing off the bound protein to reuse the sensor).
HIS1K Biosensor measuring His-tagged sfGFP. Two loading and dissociation cycles with 3 regeneration cycles in between.
Unfortunately, chat didn't know how to set reference dark and light files or do live streaming of nanometer response vs time graph. I assume this is because it didn't know how to convert raw spectrophotometer counts to nm response. It also made a loud mechanical screeching noise every time it moved the instrument head around, which wasn't great.
Expert Help
Since that didn't quite work out, I had no choice but to ask dreaded customer support for help.
The entire bio hardware and lab automation industry is known for extremely walled off and proprietary software. It's like Apple, if there were a hundred different Apples and instead of the most beautiful product design they did the opposite.
If you call Beckman Coulter and ask them for software to run their Biomek liquid handler, they will refuse to send the software over email, website portal, or any method over the internet. They will instead send a field engineer with a laptop with the software pre-installed and sell you both the laptop and consulting services for around $27,000.
Luckily, Sartorius customer support was far nicer and gave me a copy of the software for free, since they were no longer supporting the ForteBio line of instruments.
Ugly but functional (and free!) BLI data acquisition software from Sartorius
Even better, they cc'd a field application specialist (forward deployed engineer) to help out, since normally these machines are installed by expert technicians. "Robert" from Sartorius came by and was like ye it's not supposed to make that noise, and I was like ye I have no idea what I'm doing. So Robert invited me to the Sartorius showroom in Fremont for a demo (and even drove me there from SF). We entered this large, white, beautifully stocked lab with tons of gleaming, strange equipment, and I had the vague feeling of being shown around a car dealership (complete with the same general level of prices).
The Sartorius Demo Lab. Not only did they have BLI machines in 1, 2, 4, 8, 16, and 96 channel configurations, they also had bioreactors, surface plasmon resonance machines, robotic cell sorters, and more!
We spent about 8 hours on a binding kinetics protocol (human IgG + biotinylated protein A) and a quantitation experiment with some sfGFP I brought. Robert showed me everything from prepping the machine (we used an Octet R8e), to preparing the reagents and dilution series, to running the data analysis software and fitting data to theoretical curves.
Plate definition: B is assay buffer (phosphate-buffered saline solution), L is loading (protein A), the purple is the analyte (human IgG). R is acidic pH 1.5 10 mM glycine for regeneration, and N is neutralization, which is the same as assay buffer. To remove human IgG from the sensor and reuse, we do 3 regeneration cycles by quickly dipping the sensors in R and then N for 5 seconds each.
There were so many things to know, including:
It was honestly the most wonderful sales experience I ever had (we even got lunch at a nearby noodle place).
Here are some pretty curves we got at the end.
SA sensor for biotinylated protein A + human IgG. The first upward curve from 60-120 seconds is the loading of the protein A onto the sensor. This is followed by 3 regeneration cycles and a baseline soak. 210-390 is the association phase, when we dip in the human IgG solution. 390-570 is dissociation, where the bound IgG is soaked in fresh assay buffer and slowly leaves the sensor.
HIS1K sensor for His-tagged sfGFP. We didn't do regeneration and only tested loading, since it was just one protein.
Afterward, Robert gave me about $3,000 worth of biosensors, well plates, and sample reagents to troubleshoot my old machine back at the lab.
I imagine this is how children feel on Halloween
Running the Machine
Armed with new software and goodies, I tried replicating the biotinylated protein A + human IgG assay. The machine moved much more smoothly and didn't make wild noises.
There was one more problem. The machine needed initial calibration with dark (lamp off) and light (sensors in baseline) reference measurements. To set these, you have to open the software in diagnostic mode, which entails holding Ctrl + Shift while clicking the shortcut (Chat figured this out but got the timing wrong, telling me to click and then hit Ctrl+Shift on startup).
This opens diagnostic mode, which has absolutely ZERO documentation in the user guide or online (either because the machine is too old or this tacit knowledge is typically only known by a field application specialist).
What do I even do here
After a ton of back and forth with chat and hitting random buttons to observe their effects, I eventually figured out what most of the things did.
I placed two sensor tips in the sensor tray and hydrated them for 20 minutes. I noticed they were in the wrong position, and moved them to the correct positions, which didn't have assay buffer under them. This will be important in a second.
Prepping buffers, reagents, and soaking the two sensors on the right in assay buffer. There is another one of the black Greiner 655209 well plates under the green holder in the sensor tray. However, not all of these wells are filled, and the ones the protocol used to calibrate integration time were empty.
I loaded up the .fmf method file for the protocol, checked everything was good, and hit GO. The machine started the protocol, picked up the sensors, but the second it went to the first well, the sensor failed with errors (CCD signal too low) and the spectrophotometers showed zero. This was weird, because in diagnostic mode the sensors were working fine.
Data error on sensors A and B
At first I thought the calibration was off, and tried resetting the references, trying different channels, setting different INI file configs. I tried changing the integration time to 3 ms. But I noticed on assay start, it would automatically revert to ~1 ms! It turned out the sensors sitting in empty wells reflected more light than if they were sitting in assay buffer (which they were supposed to I guess) and this causes auto gain to automatically decrease integration time at the beginning of each protocol.
Notice in Instrument Status on the bottom left Sensor 1: Integration Time is set to 1.65 ms and Sensor 2 to 1.78 ms
I pipetted assay buffer under the sensors in the tray, and the sensors started working!
Biotinylated Protein A + human IgG. Red is control (assay buffer) and blue is human IgG. The nanometer response is inverted for some reason
I also tried regenerating the sensor tip but not very successfully, since at 370 seconds the curves don't drop back to 0. Here are the final curves:
Biotinylated Protein A + human IgG, two cycles + regeneration. Red is control (assay buffer) and blue is human IgG. The nanometer response is inverted for some reason
I asked chat to calculate the , , and from the data, and is apparently 59 times greater than published Octet values (indicating weaker binding), but close to another published study with porous silica.
Overall it took about $1000 and 1 month of part-time work to get from nothing to .
Limitations
Conclusion
What were the LLMs good at today?
What could have made them more helpful?
Even though they are underwhelming at present, I expect LLMs to become far more useful and powerful in the wetlab in the coming year. There are many startups like Transfyr and C5R that offer to sell multimodal wetlab data to the labs. LLMs can already (slowly) drive robot arms. Just like how LLMs could barely do math a year or so ago and are now superhuman, with the correct data and training environments the same should happen for biology.